Title :
Validation of vortex flowmeters
Author :
Clarke, D.W. ; Ghaoud, T.
Author_Institution :
Dept. of Eng. Sci., Oxford Univ., UK
Volume :
13
Issue :
5
fYear :
2002
Firstpage :
237
Lastpage :
241
Abstract :
A vortex-shedding flow meter produces an approximately sinewave signal whose average frequency is proportional to volumetric flow rate. The frequency of vortex shedding varies over a large range (up to 1:100), depending on the size of the flow meter. The conventional frequency-estimation method-a zero-crossing algorithm-fails at low flows, where the signal:noise ratio is poor. The paper describes a dual phase-locked loop structure that tracks the frequency over the flowmeter´s full range with improved accuracy. The lock indicators associated with each PLL provide the SEVA metrics.
Keywords :
flowmeters; phase locked loops; transmitters; SEVA metrics; approximately sinewave signal; average frequency; dual phase-locked loop structure; validation; volumetric flow rate; vortex-shedding flow meter;