    The trans-cardiac conductance method for on-line measurement of left ventricular volume: assessment of parallel conductance offset volume

  • Author

    Staal, Eva M. ; Steendijk, Paul ; Baan, Jan

  • Author_Institution
    Dept. of Cardiology, Leiden Univ. Med. Center, Netherlands
  • Volume
    50
  • Issue
    2
  • fYear
    2003
  • Firstpage
    234
  • Lastpage
    240
  • Abstract
    The trans-cardiac conductance (TCC) method provides on-line left ventricular (LV) volume signals by determining the electrical conductance of blood in the LV by means of central venous and epithoracic electrodes. Conductive structures outside the LV blood pool cause a "parallel conductance" offset term (Vp) that can be determined by bolus injections of hypertonic saline in the pulmonary artery (Vpsaline), which cause a transient increase in blood conductivity. This study in anesthetized sheep evaluates the accuracy of the saline calibration method and the variabilities of Vp between animals, between hemodynamic conditions and during the cardiac cycle. The conventional intra-cardiac conductance catheter method was used to obtain independent estimates of Vp by the zero-volume method (Vpzerovolume). Mean baseline Vpsaline and Vpzerovolume were 104 ± 6 ml and 106 ± 6 ml, respectively. Bland-Altman analysis showed a small nonsignificant bias (-2.5 ml) and narrow limits of agreement (4.6 ml). Vp was not significantly different between hemodynamic conditions (baseline, dobutamine, volume load, propranolol), but had a substantial interanimal variability (IAV) (38%). Average variations during the cardiac cycle were <10% of mean Vp. We conclude that the saline method can be applied to determine Vp for TCC. IAV is substantial, so that Vp must be determined in each animal, but within-animal variability is relatively small.
  • Keywords
    bioelectric phenomena; blood; calibration; cardiology; electric admittance measurement; patient monitoring; volume measurement; Bland-Altman analysis; blood conductivity increase; cardiac cycle; dobutamine; hemodynamic conditions; left ventricular volume; on-line measurement; parallel conductance offset volume; propranolol; transcardiac conductance method; volume load; within-animal variability; zero-volume method; Animals; Biomedical imaging; Blood; Cardiac disease; Catheters; Electrodes; Hemodynamics; Medical diagnostic imaging; Monitoring; Volume measurement; Animals; Cardiac Volume; Cardiography, Impedance; Dobutamine; Electric Conductivity; Gelatin; Propranolol; Reproducibility of Results; Sensitivity and Specificity; Sheep; Sodium Chloride; Statistics as Topic; Stroke Volume; Succinates; Ventricular Function, Left;
  • fLanguage
    English
  • Journal_Title
    Biomedical Engineering, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9294
  • Type

    jour

  • DOI
    10.1109/TBME.2002.807646
  • Filename
    1185147