• DocumentCode
    1159300
  • Title

    Immunity tests of implantable cardiac pacemaker against CW and pulsed ELF fields: experimental and numerical results

  • Author

    Augello, Antonio ; Chiara, Giuseppe Della ; Primiani, Valter Mariani ; Moglie, Franco

  • Author_Institution
    Dipt. di Electrornagnetismo e Bioingegneria, Univ. Politecnica della Marche, Ancona
  • Volume
    48
  • Issue
    3
  • fYear
    2006
  • Firstpage
    502
  • Lastpage
    515
  • Abstract
    The aim of the present work is to investigate pacemaker (PMK) immunity against high-level extremely low frequency (typically 50 Hz) magnetic fields that can be found in industrial environment where high power machines operate. To that purpose, a test bench has been set up using a Helmholtz coil for producing extremely low frequency magnetic fields and a trunk simulator rightly fed by cardiac signals. A widely used PMK was tested, and results have been reported, under different operating conditions, for both continuous waves and variously pulsed excitations. A numerical tool has also been developed to analyze the realized test bench, based on a proper discretization of a human trunk simulation and on the construction of an equivalent three-dimensional (3-D) network. Numerical results have showed a good agreement with the experimental results
  • Keywords
    Helmholtz equations; biological effects of fields; cardiology; immunity testing; magnetic fields; pacemakers; CW ELF fields; Helmholtz coil; cardiac signals; equivalent three-dimensional network; extremely low frequency magnetic fields; human trunk simulation; immunity tests; implantable cardiac pacemaker; pulsed ELF fields; test bench; Coils; Frequency; Geophysical measurement techniques; Ground penetrating radar; Humans; Immune system; Magnetic analysis; Magnetic fields; Pacemakers; Testing; ELF fields; Helmholtz coil; immunity tests; pacemaker (PMK);
  • fLanguage
    English
  • Journal_Title
    Electromagnetic Compatibility, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9375
  • Type

    jour

  • DOI
    10.1109/TEMC.2006.877783
  • Filename
    1677784