DocumentCode :
1164075
Title :
Frequency Dependence of Microwave Radiometer Calibration
Author :
Goodberlet, Mark A. ; Popstefanija, Ivan
Author_Institution :
ProSensing Inc., Amherst, MA
Volume :
6
Issue :
2
fYear :
2009
fDate :
4/1/2009 12:00:00 AM
Firstpage :
337
Lastpage :
338
Abstract :
A consequence of impedance mismatch for components located in the front-end of a microwave radiometer is that the radiometer´s brightness temperature calibration becomes frequency-dependent. This dependence can limit the bandwidth over which a particular calibration is valid and place restrictions on local oscillator drift.
Keywords :
calibration; geophysical equipment; microwave measurement; radiometers; radiometry; brightness temperature; front-end; impedance mismatch; local oscillator drift; microwave radiometer calibration; Calibration; microwave radiometry; microwave receivers;
fLanguage :
English