• DocumentCode
    1167657
  • Title

    Detective quantum efficiency [DQE(0)] of CZT semiconductor detectors for digital radiography

  • Author

    Giakos, G.C. ; Suryanarayanan, S. ; Guntupalli, R. ; Odogba, J. ; Shah, N. ; Vedantham, S. ; Chowdhury, S. ; Mehta, K. ; Sumrain, S. ; Patnekar, N. ; Moholkar, A. ; Kumar, V. ; Endorf, R.E.

  • Author_Institution
    Dept. of Electr. & Comput. Eng., Univ. of Akron, OH, USA
  • Volume
    53
  • Issue
    6
  • fYear
    2004
  • Firstpage
    1479
  • Lastpage
    1484
  • Abstract
    In this paper, the detective quantum efficiency (DQE) of cadmium zinc telluride (CZT) detector samples for digital radiography has been measured. Specifically, this study is aimed at investigating the zero-frequency DQE(0) under different X-ray tube and detector parameters. The experimental results of this study indicate that the DQE(0) of the CZT samples is strongly dependent upon the irradiation geometry. This is attributed to the incomplete charge collection process, which can be further improved by controlling the purity of the samples and the contact type.
  • Keywords
    X-ray tubes; cadmium; radiography; semiconductor counters; tellurium; zinc; CdZnTe; X-ray tube; cadmium zinc telluride; charge collection process; detective quantum efficiency; detector parameters; digital radiography; irradiation geometry; semiconductor detectors; zero-frequency DQE(O); Biomedical imaging; Cadmium compounds; Detectors; Image sensors; Photoconductivity; Radiography; Sensor arrays; X-ray detection; X-ray imaging; Zinc compounds; 65; Detective quantum efficiency; digital radiography; semiconductor detectors;
  • fLanguage
    English
  • Journal_Title
    Instrumentation and Measurement, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9456
  • Type

    jour

  • DOI
    10.1109/TIM.2004.834590
  • Filename
    1360085