An axial gas ionization chamber has been fabricated for use as a DeltaE detector in heavy ion induced nuclear reactions. Different operating parameters such as gas type, pressure, anode voltage, and anode structures have been optimized. The transparency of the anode structure is observed to play an important role in improving the energy resolution of the detector
Keywords :
heavy ion-nucleus reactions; ionisation chambers; radiation detection; spontaneous fission; anode structures; anode voltage; axial field; axial gas ionization chamber; energy resolution; gas pressure; heavy ion induced nuclear reactions; radiation detection; spontaneous fission fragments; Anodes; Energy resolution; Gas detectors; Ionization chambers; Plasma measurements; Radiation detectors; Silicon; Solid state circuits; Timing; Voltage; Anode structure; axial field; gas ionization chamber;