DocumentCode :
1169074
Title :
A miniature hybrid reflection type optical sensor for measurement of hemoglobin content and oxygen saturation of whole blood
Author :
Takatini, S. ; Noda, Hiroyuki ; Takano, Hisateru ; Akutsu, Tetsuzo
Author_Institution :
Nat. Cardiovascular Center, Res. Inst., Osaka, Japan
Volume :
35
Issue :
3
fYear :
1988
fDate :
3/1/1988 12:00:00 AM
Firstpage :
187
Lastpage :
198
Abstract :
A reflection-type hybrid optical sensor was designed for continuous measurement of both hemoglobin content ((Hb)) and oxygen saturation (OS) of whole blood. The sensor was based on a theoretical model using the photon diffusion theory. The prototype sensor consists of a light emitting diode (LED) and photodiode chips mounted on a single substrate and cased in a TO-5 can with its surface covered with clear epoxy resin. The effect of hematocrit variation was eliminated. It was test in goat, calf, and human whole blood over hematocrit range of 20-45%. The modified ratio method reduced the effect of hematocrit variation of hemoglobin OS measurement with improvement in standard deviation of errors from about 7.0 to 2.0%. The standard errors in the estimation of hemoglobin content ranged from 1.5 to 0.5 gm%. The oxygen contents of whole blood derived from the prototype sensor correlated well (r=0.997) with those analyzed using a hemoximeter. The prototype sensors have been used for continuous measurement of oxygen delivery to, and oxygen consumption by, peripheral tissues in artificial heart animals for a duration of 40 days, demonstrating satisfactory performance.
Keywords :
biological techniques and instruments; blood; oxygen; proteins; O/sub 2/ saturation measurement; artificial heart animals; calf; clear epoxy resin; goat; hematocrit; hemoglobin content; human; light emitting diode; miniature hybrid reflection type optical sensor; peripheral tissues; photodiode chips; photon diffusion theory; standard errors; whole blood; Blood; Epoxy resins; Light emitting diodes; Optical design; Optical reflection; Optical sensors; Optoelectronic and photonic sensors; Photodiodes; Prototypes; Semiconductor device measurement; Animals; Erythrocyte Indices; Hemoglobinometry; Humans; Models, Biological; Optics; Oximetry;
fLanguage :
English
Journal_Title :
Biomedical Engineering, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9294
Type :
jour
DOI :
10.1109/10.1361
Filename :
1361
Link To Document :
بازگشت