Title
Optimization of the use of small inexpensive side-window photomultiplier tubes for scintillator readout
Author
Weisenberger, A.G. ; Cheung, F. ; Kross, B. ; Majewski, S. ; Wojcik, R. ; Zorn, C.
Author_Institution
Continuous Electron Beam Accel. Facility, Newport News, VA, USA
Volume
41
Issue
4
fYear
1994
fDate
8/1/1994 12:00:00 AM
Firstpage
708
Lastpage
712
Abstract
The 931B is a small inexpensive side-window photomultiplier tube (PMT) which has recently been found to be useful for the readout of scintillators. We present the results of a study to obtain improved performance from the PMT by facilitating the transport of light to the reflection type photocathode. Particular attention is given for the use of these PMTs for the readout of light guides, waveshifting and scintillating fibers. A CCD camera was positioned behind a glass envelope to simulate the photocathode to aid in quantifying the optimum combination of optical coupling and aiming of the fiber. We will also report on tests comparing a similar side-window PMT, the Thorn EMI model 9781B to the 931B. Tests on the operation of the side-window PMT at several orientations in a uniform magnetic field will also be reported. Comparisons between the side-window PMT and a RCA 31000M head-on tube will also be reported
Keywords
photomultipliers; scintillation counters; CCD camera; Hamamatsu 931B; RCA 31000M; Thorn EMI 9781B; light guides; scintillating fibers; scintillator readout; side-window photomultiplier tubes; waveshifting fibers; Cathodes; Charge coupled devices; Charge-coupled image sensors; Electromagnetic interference; Glass; Magnetic fields; Optical coupling; Optical fiber testing; Optical reflection; Photomultipliers;
fLanguage
English
Journal_Title
Nuclear Science, IEEE Transactions on
Publisher
ieee
ISSN
0018-9499
Type
jour
DOI
10.1109/23.322793
Filename
322793
Link To Document