• DocumentCode
    1176603
  • Title

    Vertical focusing device utilizing dielectrophoretic force and its application on microflow cytometer

  • Author

    Lin, Che-Hsin ; Lee, Gwo-Bin ; Fu, Lung-Ming ; Hwey, Bao-Herng

  • Author_Institution
    Dept. of Mech. & Electro-Mech. Eng., Nat. Sun Yat-Sen Univ., Kaohsiung, Taiwan
  • Volume
    13
  • Issue
    6
  • fYear
    2004
  • Firstpage
    923
  • Lastpage
    932
  • Abstract
    Focusing of particles/cells in the vertical direction inside a micromachined flow cytometer is a critical issue while using an embedded optical detection system aligned with microchannels. Even if the particles/cells have been focused centrally in the horizontal direction using coplanar sheath flows, appreciable errors may still arise if they are randomly distributed in the vertical direction. This work presents a vertical focusing device utilizing dielectrophoretic (DEP) forces and its application on micromachined flow cytometer. A pair of parallel microelectrodes is deposited on the upper and bottom surface of the microfluidic channel to drive particles/cells into the vertical center of the sample flow. This new microfluidic device is capable of three-dimensional (3-D) focusing of microparticles/cells and thus improves the uniformity of the optical detection signals. This 3-D focusing feature of the sample flow is realized utilizing the combination of dielectrophoretic and hydrodynamic forces. Initially, two sheath flows are used to focus the sample flow horizontally by means of hydrodynamic forces, and then two embedded planar electrodes apply negative DEP forces to focus the particles/cells vertically. A new micromachined flow cytometer integrated with an embedded optical detection mechanism is then demonstrated. Numerical simulation is used to analyze the operation conditions and the dimension of the microelectrodes for DEP manipulation. The dynamic trace of the moving particles/cells within a flow stream under the DEP manipulation is calculated numerically. Micro polystyrene beads and diluted human red blood cells (RBC) are used to test the performance of the proposed device. The experimental results confirm the suitability of the proposed device for applications requiring precise counting of particles or cells. Experimental data indicates the proposed method can provide more stable signals over the other types of micromachined flow cytometers that were previously reported.
  • Keywords
    biological techniques; cellular biophysics; electrophoresis; fibre optic sensors; microelectrodes; microfluidics; 3D focusing; dielectrophoretic force; hydro-dynamic force; micro polystyrene beads; microelectromechanical systems; microflow cytometer; microfluidic channel; numerical simulation; optical detection signals; parallel microelectrodes; planar electrodes; red blood cells; vertical focusing device; Dielectrophoresis; Electrodes; Hydrodynamics; Image motion analysis; Microchannel; Microelectrodes; Microfluidics; Optical detectors; Optical devices; Signal detection; -D; 65; Dielectrophoretic force; MEMS; flow cytometer; focusing; hydrodynamic force; microelectromechanical systems; three-dimensional;
  • fLanguage
    English
  • Journal_Title
    Microelectromechanical Systems, Journal of
  • Publisher
    ieee
  • ISSN
    1057-7157
  • Type

    jour

  • DOI
    10.1109/JMEMS.2004.838352
  • Filename
    1364050