Title :
Response of silicon detector for high energy X-ray computed tomography
Author :
Miyai, H. ; Kawasaki, S. ; Kitaguchi, H. ; Izumi, S.
Author_Institution :
Energy Res. Lab., Hitachi Ltd., Ibaraki, Japan
fDate :
8/1/1994 12:00:00 AM
Abstract :
The suitability of an Si-SSD as a detector for high energy X-ray CT has been investigated. Two types of Si-SSDs which can take the place of ordinary scintillation detectors were proposed. Their predicted responses were compared with that of a CWO scintillation detector of almost the same dimensions by using the EGS4 Monte Carlo simulation code. The sensitivities of the sole-silicon detector and the metal-sandwiched detector were about ten times higher than that of the CWO scintillation detector at X-ray energies above 1 MeV. The metal-sandwiched detector was judged more useful than the sole Si-SSD when the width of the X-ray beam was less than about 0.2 mm. The silicon detector for high energy X-ray CT was fabricated and actual measurements of the depletion layer width and the electronic noise contribution were made
Keywords :
Monte Carlo methods; X-ray detection and measurement; biomedical equipment; computerised tomography; semiconductor counters; 1 MeV; CWO scintillation detector; EGS4; Si; Si detector; depletion layer width; high energy X-ray computed tomography; noise; response; sensitivities; Attenuation; Computed tomography; Object detection; Scintillation counters; Silicon; Solid scintillation detectors; Spatial resolution; X-ray detection; X-ray detectors; X-ray imaging;
Journal_Title :
Nuclear Science, IEEE Transactions on