• DocumentCode
    1176848
  • Title

    Low energy x-ray response of Ge detectors with amorphous Ge entrance contacts

  • Author

    Luke, P.N. ; Rossington, C.S. ; Wesela, M.F.

  • Author_Institution
    Div. of Eng., California Univ., Berkeley, CA, USA
  • Volume
    41
  • Issue
    4
  • fYear
    1994
  • fDate
    8/1/1994 12:00:00 AM
  • Firstpage
    1074
  • Lastpage
    1079
  • Abstract
    The low energy X-ray response of Ge detectors with amorphous Ge entrance contacts has been evaluated. The spectral background due to near contact incomplete charge collection was found to consist of two components: a low level component which is insensitive to applied voltage and a high level step-like component which is voltage dependent. At high operating voltages, the high level component can be completely suppressed, resulting in background levels which are much lower than those previously observed using Ge detectors with Pd surface barrier or B ion implanted contacts, and which also compare favorably to those obtained with Si(Li) X-ray detectors. The response of these detectors to 55Fe and 1.77 keV X-rays is shown. A qualitative explanation of the origins of the observed background components is presented
  • Keywords
    X-ray detection and measurement; semiconductor counters; Ge; Ge detectors; amorphous Ge entrance contacts; background components; high level component; low energy X-ray response; near contact incomplete charge collection; spectral background; Amorphous materials; Contacts; Electromagnetic wave absorption; Gold; Laboratories; Power engineering and energy; Spectroscopy; Voltage; X-ray detection; X-ray detectors;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/23.322861
  • Filename
    322861