  • Abstract
    New experimental data is presented which displays 137Cs resolution of both single Si(Li) devices and a detector stack 2 cm in height as a function of temperature (85 K⩽T⩽245 K). We also discuss variations in photopeak shape which indicate that detector charge collection may be temperature dependent over the range of interest
  • Keywords
    gamma-ray detection and measurement; semiconductor counters; 85 to 245 K; Si(Li) detector stack; Si:Li; charge collection; photopeak shape; temperature dependence; Cooling; Detectors; Displays; Laboratories; Leakage current; Temperature control; Temperature dependence; Temperature distribution; Temperature measurement; Testing;