Title :
Modeling the scatter response function in inhomogeneous scattering media for SPECT
Author :
Frey, E.C. ; Tsui, B.M.W.
Author_Institution :
Dept. of Biomed. Eng., North Carolina Univ., Chapel Hill, NC, USA
fDate :
8/1/1994 12:00:00 AM
Abstract :
The authors have previously developed a method for accurately and rapidly modeling the scatter response function in uniform media. Here they studied the extension of this method to nonuniform attenuators. Their approach was to use the water-equivalent source depth, i.e., the integral of the attenuation coefficient through the nonuniform object from the source to the surface divided by the attenuation coefficient of water. The authors have investigated the accuracy of 3 different methods based on this approximation using Monte Carlo (MC) simulation methods. Line sources were placed at the same water equivalent depth in slabs composed of water, bone, and lung. The authors observed that, for perfect collimation, the scatter response functions (SRFs) obtained for these slabs are comparable. To determine whether attenuation coefficients alone are sufficient to estimate the SRF or whether one must know the elemental composition, the authors simulated SRFs for sources in bone and lung as well as bone- and lung-equivalent materials. The bone- and lung-equivalent materials have the same attenuation coefficient as bone and lung, respectively, but the same elemental composition as water. For the lung-equivalent material, the SRFs were essentially equivalent to those from lung; for the bone-equivalent material, the SRFs were closer to those from water. The authors have also placed voids of various sizes in slab phantoms while keeping the same water-equivalent source depth. For these simple geometries 2 of the methods based on the effective depth were adequate for predicting the SRF. Finally, the authors have simulated the SRF for a line source in a realistic thorax phantom using several methods based on geometric and water-equivalent distances. The results indicate that one of the methods gives reasonably good agreement with direct MC simulations in terms of predicting the magnitude and shape of the SRFs and projection data for a complex distribution simulating the heart. This method was not the same as the one that gave good agreement in the case of the simpler phantoms
Keywords :
computerised tomography; gamma-ray scattering; modelling; radioisotope scanning and imaging; Monte Carlo simulation methods; SPECT; bone; heart; inhomogeneous scattering media; line source; lung; medical diagnostic imaging; nonuniform attenuators; nuclear medicine; scatter response function modeling; single photon emission computerised tomography; thorax phantom; uniform media; water; water attenuation coefficient; Attenuation; Attenuators; Bones; Composite materials; Imaging phantoms; Lungs; Predictive models; Scattering; Slabs; Water resources;
Journal_Title :
Nuclear Science, IEEE Transactions on