Title :
Calculation and Digital Display of Whole Blood Oxygen Saturation by Analog Techniques
Author :
Sutterer, William F. ; Krovetz, L.Jerome
Author_Institution :
Waters Company, Rochester, Minn. 55901.
fDate :
4/1/1969 12:00:00 AM
Abstract :
An electronic analog system has been developed which performs the double scale calculation from the Wood oximeter and gives a single output as a log ratio and also gives a digital display of oxygen saturation, with simultaneous recording of these parameters if desired. This system, built of commercially available analog components, consists of operational amplifiers to amplify the output of the red and infrared photocells, logarithmic devices to calculate the optical density as sensed by these two photocells, a division circuit to calculate the log ratio, a scaling circuit to facilitate digital display in percent oxygen saturation, and a digital voltmeter.
Keywords :
Blood; Circuits; Digital recording; Displays; Operational amplifiers; Optical amplifiers; Optical recording; Oxygen; Semiconductor optical amplifiers; Stimulated emission; Blood Gas Analysis; Computers; Computers, Analog; Humans; Mathematics; Methods; Oximetry; Oxygen Consumption; Spectrophotometry;
Journal_Title :
Biomedical Engineering, IEEE Transactions on
DOI :
10.1109/TBME.1969.4502623