DocumentCode :
1183217
Title :
Nonlinearity of Cardiac Pacemaker Electrodes
Author :
Jaron, Dov ; Briller, Stanley A. ; Schwan, Herman P. ; Geselowitz, David B.
Author_Institution :
Surgical Research Laboratory, Maimonides Medical Center, Brooklyn, N. Y. 11219.
Issue :
2
fYear :
1969
fDate :
4/1/1969 12:00:00 AM
Firstpage :
132
Lastpage :
138
Abstract :
Nonlinear properties of electrodes immersed in saline were studied in the time domain and frequency domain. Three electrode materials were tested: platinum-iridium, Elgiloy, and stainless steel. For time domain measurements, current pulses of 1.0 to 2.0 ms duration and varying amplitudes were applied to the electrodes. Frequency domain measurements of polarization impedance were made with an impedance bridge, utilizing sinusoidal currents of varying amplitudes at frequencies between 0.1 and 10 kHz.
Keywords :
Building materials; Current measurement; Electrodes; Frequency domain analysis; Impedance; Materials testing; Pacemakers; Pulse measurements; Steel; Time measurement; Alloys; Electrodes; Equipment and Supplies; Iridium; Methods; Pacemaker, Artificial; Platinum;
fLanguage :
English
Journal_Title :
Biomedical Engineering, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9294
Type :
jour
DOI :
10.1109/TBME.1969.4502625
Filename :
4502625
Link To Document :
بازگشت