DocumentCode
1183355
Title
Design Analysis for a Solenoid Blood Pump Actuator
Author
Fuller, Jack W. ; Armijo, Larry
Author_Institution
Bio-Medical Laboratory, Rice University, Houston, Tex.
Issue
3
fYear
1969
fDate
7/1/1969 12:00:00 AM
Firstpage
184
Lastpage
191
Abstract
An analysis of a solenoid suitable for direct actuation of a physiological blood pump is carried out in this paper. A general theoretical study of the electromechanical energy conversion process in the solenoid is presented, as well as a discussion of the characteristics of the magnetic circuit configuration. A computer program is described which is based on these results and is used to optimize the solenoid design with respect to conversion efficiency.
Keywords
Actuators; Blood; Design optimization; Energy conversion; In vitro; In vivo; Laboratories; Magnetic analysis; Magnetic circuits; Solenoids; Computers; Electronics, Medical; Heart, Artificial; Humans; Mathematics;
fLanguage
English