DocumentCode
1184177
Title
An Integrated-Circuit Approach to Extracellular Microelectrodes
Author
Wise, Kensall D. ; Angell, James B. ; Starr, Arnold
Author_Institution
Department of Electrical Engineering, Stanford University, Stanford, Calif.
Issue
3
fYear
1970
fDate
7/1/1970 12:00:00 AM
Firstpage
238
Lastpage
247
Abstract
This paper describes a new multielectrode microprobe which utilizes integrated-circuit fabrication techniques to overcome many of the problems associated with conventional microelectrodes. The probe structure consists of an array of gold electrodes which are supported on a silicon carrier and which project beyond the carrier for a distance of about 50 ¿ to allow a close approach to active neurons. These electrodes are covered with a thin (0.4-¿) layer of silicon dioxide which is selectively removed at the electrode tips using high-resolution photoengraving techniques to define the recording areas precisely. The processing sequence described permits any two-dimensional electrode array to be realized. Interelectrode spacings can be accurately controlled in the range from 10 to 20 ¿ or larger, and electrode-tip diameters can be as small as 2 ¿.
Keywords
Electrodes; Extracellular; Fabrication; Gold; Insulation; Microelectrodes; Neurons; Probes; Silicon compounds; Testing; Animals; Biomedical Engineering; Cats; Electrodes; Electronics, Medical; Neurophysiology;
fLanguage
English
Journal_Title
Biomedical Engineering, IEEE Transactions on
Publisher
ieee
ISSN
0018-9294
Type
jour
DOI
10.1109/TBME.1970.4502738
Filename
4502738
Link To Document