Title :
Measurement of spatial cross sections of ultrasound pressure fields by optical scanning means
Author :
Klann, Martin ; Koch, Christian
Author_Institution :
Dept. of Sound, Phys. Tech. Bundesanstalt, Braunschweig, Germany
Abstract :
The measurement of spatial cross sections of ultrasound pressure fields is an essential element of exposimetry of ultrasonic medical equipment. An optical technique is presented that allows the two-dimensional (2-D) determination of ultrasound pressure using an optical multilayer hydrophone in which a laser beam with suitable wavelength is partially reflected from a dielectric optical multilayer system. By detecting the change in reflectivity of the multilayer coating induced by the incident ultrasound, the pressure time waveform can be determined. A 2-D data acquisition covering an area of at least 15 mm /spl times/ 5 mm was realized by two complementary approaches. A serial detection scheme was set up by scanning the sensing point across the area of interest by a micromechanically engineered scanning mirror and acquiring pressure time waveforms sequentially and pointwise. This allows the measurement of repeating ultrasonic waveforms with a spatial resolution of better than 70 /spl mu/m and a minimal detectable pressure of 50 kPa (bandwidth: 50 MHz) in a few seconds. In an alternative approach exploiting the parallel processing capabilities of a charge-coupled devices (CCD) camera chip, the probe was strobe-illuminated by a large-diameter collimated beam of a pulsed laser diode. The 2-D pressure distribution at a particular moment was derived from captured reflectivity distributions with a spatial resolution of 75 /spl mu/m. By successive delaying of the laser pulse with respect to the ultrasound pulse, the complete 2-D pulse waveform was acquired with high spatial resolution. Measurement results on ultrasound fields from plane and focusing transducers are presented and compared to simulation results. Individual advantages and drawbacks of both approaches are discussed. A combined setup merging both detection schemes into a single device was developed and represents a milestone on the way toward constructing an optical ultrasound measuring camera.
Keywords :
CCD image sensors; biomedical measurement; biomedical transducers; biomedical ultrasonics; dielectric materials; hydrophones; laser applications in medicine; micromechanical devices; micromirrors; optical multilayers; optical sensors; pressure measurement; reflectivity; 15 mm; 2D data acquisition; 2D pressure distribution; 5 mm; 50 MHz; 50 kPa; camera chip; charge-coupled devices; dielectric optical multilayer; exposimetry; focusing transducers; large-diameter collimated beam; laser beam; laser pulse delay; micromechanically engineered scanning mirror; multilayer coating; optical multilayer hydrophone; optical scanning; optical technique; optical ultrasound measuring camera; parallel processing; partial reflection; plane transducers; pressure time waveform; pressure time waveforms; pulsed laser diode; reflectivity; repeating ultrasonic waveforms; serial detection scheme; spatial cross sections; spatial resolution; strobe-illuminated probe; two-dimensional ultrasound pressure; ultrasonic medical equipment; ultrasound pressure fields; ultrasound pulse; Biomedical optical imaging; Laser beams; Optical pulses; Optical sensors; Optical superlattices; Pressure measurement; Reflectivity; Spatial resolution; Ultrasonic imaging; Ultrasonic variables measurement;
Journal_Title :
Ultrasonics, Ferroelectrics, and Frequency Control, IEEE Transactions on
DOI :
10.1109/TUFFC.2005.1516027