DocumentCode
1184782
Title
A Comparison of Tonographic-Trace Simulations Achieved by Two Systems: Nonlinear Elastic and Linear Viscoelastic
Author
Swigert, Charles J. ; Shepherd, Marvin H. ; Lyon, Catherine S. ; McEwen, William K.
Author_Institution
TRW Systems Group, Redondo Beach, Calif. 90278.
Issue
2
fYear
1971
fDate
3/1/1971 12:00:00 AM
Firstpage
97
Lastpage
103
Abstract
Twenty normal clinical tonographic traces were averaged and compared at 10-s intervals by means of two systems: 1) the decay curves of a linear viscoelastic mathematical model having a single degree of freedom, and 2) nonlinear elastic models, including the classical (or current) Grant model. The linear viscoelastic model provided a considerably better match than the other models. So far as degree of match and facility of outflow values were concerned, the study showed the effect on the models of various investigators´ formulations of ocular rigidity, corneal indentation, and tonometer calibration. In precise work facility of outflow values should be reported only if a measure of the degree of fit to the tonogram is given.
Keywords
Calibration; Elasticity; Health and safety; Humans; Mathematical model; Mechanical factors; Performance analysis; Public healthcare; Springs; Viscosity; Elasticity; Mathematics; Models, Biological; Tonometry, Ocular;
fLanguage
English
Journal_Title
Biomedical Engineering, IEEE Transactions on
Publisher
ieee
ISSN
0018-9294
Type
jour
DOI
10.1109/TBME.1971.4502809
Filename
4502809
Link To Document