• DocumentCode
    1185245
  • Title

    An Optimally Controlled Respirator

  • Author

    Mitamura, Yoshinori ; Mikami, Tomohisa ; Sugawara, Hiromi ; Yoshimoto, Chiyoshi

  • Author_Institution
    Department of Medical Electronics, Research Institute of Applied Electricity, Hokkaido University, Sapporo, Japan.
  • Issue
    5
  • fYear
    1971
  • Firstpage
    330
  • Lastpage
    338
  • Abstract
    An optimally controlled respirator was developed. It has three main features: 1) ventilation is controlled by the patient´s metabolic rate from continuously measured C02 output, 2) physiologic dead space approximated as a linear function of tidal volume is used to estimate alveolar ventilation, and 3) respiratory rate is computed to minimize ventilatory work.
  • Keywords
    Automatic control; Control systems; Dogs; Extraterrestrial measurements; Linear approximation; Medical control systems; Optimal control; Surgery; Ventilation; Volume measurement; Animals; Basal Metabolism; Blood Gas Analysis; Dogs; Humans; Respiration, Artificial; Respiratory Insufficiency; Ventilators, Mechanical;
  • fLanguage
    English
  • Journal_Title
    Biomedical Engineering, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9294
  • Type

    jour

  • DOI
    10.1109/TBME.1971.4502864
  • Filename