• DocumentCode
    1188617
  • Title

    A Gated Radiometer for Plasma Afterglow Study

  • Author

    Taft, W.C. ; Stotz, K.C. ; Holt, E.H.

  • Author_Institution
    Rensselaer Polytechnic Institute, Troy, N. Y.
  • Volume
    12
  • Issue
    2
  • fYear
    1963
  • Firstpage
    90
  • Lastpage
    96
  • Abstract
    In order to monitor the electron temperature during plasma afterglow experiments, a gated microwave radiometer has been developed. The radiometer differs from the usual Dicke type of radiometer in the input switching scheme and in the fact that a good time resolution is required. The radiometer to be described has a time resolution of one microsecond and a temperature sensitivity of about ten degrees Kelvin when measuring temperatures near room temperature. The response time when used with a repetitive afterglow is 20 seconds. Measurements of electron temperature in a decaying nitrogen plasma have been made duirng the initial 300 microseconds of the afterglow.