DocumentCode :
1188815
Title :
A New Method of Measuring Gap Voltages and Parallel Resistances of Cavities
Author :
Tove, P.A. ; Jonasson, L.G.
Issue :
1
fYear :
1964
fDate :
3/1/1964 12:00:00 AM
Firstpage :
29
Lastpage :
33
Abstract :
A new method of measuring voltages across cavity gaps and deducing the cavity´s parallel resistance is discussed. The method is based on velocity modulation of a beam of monoenergetic charged particles (electrons) from a radioactive source. The change in the line spectra of an energy-selective particle detector (a semiconductor detector), when the particles pass through the gap, is used for calculating the cavity gap voltage. Combination with a measurement of the power dissipated in the cavity gives the value of the parallel resistance. A description is given of a practical arrangement together with experimental results.
Keywords :
Electrical resistance measurement; Electron beams; Energy measurement; Energy resolution; Pulse amplifiers; Pulse measurements; Q measurement; Solid scintillation detectors; Spectroscopy; Voltage;
fLanguage :
English
Journal_Title :
Instrumentation and Measurement, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9456
Type :
jour
DOI :
10.1109/TIM.1964.4313365
Filename :
4313365
Link To Document :