A Standard Attenuator and the Precise Measurement of Attenuation
Author :
Hollway, D.L. ; Kelly, F.P.
fDate :
3/1/1964 12:00:00 AM
Abstract :
An attenuator calibration system is described consisting of a waveguide-beyond-cutoff attenuator operating at 31.25 Mc and electronic equipment designed for parallel-substitution, series-substitution, and heterodyne measurements. Radio-frequency power is supplied by a crystal-controlled signal generator, stabilized so that variations in the power output do not exceed 0.001 db per hour. Measurements of a 10-db step made by the standard attenuator, and by two other methods, agree to within 0.001 db. A number of possible sources of error are examined including nonlinearity in microwave mixers.
Keywords :
Attenuation measurement; Attenuators; Australia; Calibration; Electronic equipment; Impedance; Insertion loss; Measurement standards; Reflection; Standards development;