DocumentCode :
1188921
Title :
Further Analysis of the Modulated Subcarrier Technique of Attenuation Measurement
Author :
Little, William E.
fYear :
1964
Firstpage :
71
Lastpage :
76
Abstract :
The modulated subcarrier technique of attenuation measurement offers an accurate and convenient means of calibrating rotary vane attenuators. The technique also presents the possibility of being extended to other types of attenuation measurement. This paper gives an analysis of the output of the microwave detector, assuming it is completely linear; an error analysis of the technique; and a comparison of the technique with the 30-Mc IF substitution technique of attenuation measurement.
Keywords :
Amplitude modulation; Attenuation measurement; Attenuators; Blades; Chirp modulation; Envelope detectors; Error analysis; Microwave theory and techniques; Signal detection; Voltage;
fLanguage :
English
Journal_Title :
Instrumentation and Measurement, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9456
Type :
jour
DOI :
10.1109/TIM.1964.4313377