  • Title

    Quality control of intensity modulated radiation therapy with optically stimulated luminescent films

  • Author

    Idri, K. ; Santoro, L. ; Charpiot, E. ; Herault, J. ; Costa, A. ; Aillères, N. ; Delard, R. ; Vaillé, J.R. ; Fesquet, J. ; Dusseau, L.

  • Author_Institution
    CEM2, Univ. Montpellier, France
  • Volume
    51
  • Issue
    6
  • fYear
    2004
  • Firstpage
    3638
  • Lastpage
    3641
  • Abstract
    An optically stimulated luminescent (OSL) dose mapping system is characterized and its possible application to intensity modulated radiotherapy therapy (IMRT) treatment investigated. The dose map obtained during an IMRT treatment performed on a Plexiglas phantom corroborates the results obtained with radiological films.
  • Keywords
    dosimetry; luminescent devices; optical films; phantoms; quality control; radiation monitoring; radiation therapy; dosimetry; intensity modulated radiotherapy; optically stimulated luminescent dose mapping system; optically stimulated luminescent films; plexiglas phantom; quality control; radiological films; Biomedical applications of radiation; Imaging phantoms; Intensity modulation; Luminescence; Medical treatment; Optical devices; Optical films; Optical modulation; Quality control; Stimulated emission;