• DocumentCode
    1190609
  • Title

    A Coaxial Radio-Frequency Power Standard

  • Author

    Jurkus, A.

  • Volume
    15
  • Issue
    4
  • fYear
    1966
  • Firstpage
    338
  • Lastpage
    342
  • Abstract
    A coaxial calorimeter intended for use as a reference standard of radio-frequency power is described. It is designed to operate at power levels up to 100 mW and frequencies up to 6 GHz. The calorimeter is of the twin-load type; one load being the calorimetric body while the other provides a reference level of temperature. The input connector is a General Radio type 900 precision connector. The sensitivity of the calorimeter is about 50 ¿V/mW and this has been achieved by means of a multijunction thermopile. The uncertainty in the measured power level has been estimated to vary from ±0.20 percent at low frequencies to ±0.37 percent at 6 GHz.
  • Keywords
    Coaxial components; Conductors; Connectors; Copper; Frequency estimation; Power measurement; Radio frequency; Resistors; Temperature sensors; Thermal conductivity;
  • fLanguage
    English
  • Journal_Title
    Instrumentation and Measurement, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9456
  • Type

    jour

  • DOI
    10.1109/TIM.1966.4313560
  • Filename
    4313560