• DocumentCode
    1190659
  • Title

    Calibration and Stability of a Millimeter Confocal Wavemeter

  • Author

    Zamites, C.J., Jr.

  • Volume
    15
  • Issue
    4
  • fYear
    1966
  • Firstpage
    388
  • Lastpage
    392
  • Abstract
    The construction and calibration of a confocal wavemeter for the 94 GHz region is described. Material selection was predicated upon obtaining a frequency stability of 1 part in 105 per °C over a 34°C temperature range. The system uses a 94 GHz signal generated from a crystal reference and phase-lock loop to calibrate the wavemeter. Several readout techniques are discussed. 1) the conventional dip indication, 2) the pip obtained from the resonator used in the transmission mode, and 3) the measurement of the pip phase. The latter technique permits the resolution of 100 KHz at a center frequency of 100 GHz or 1 part in 106.
  • Keywords
    Calibration; Crystalline materials; Electromagnetic measurements; Frequency measurement; Millimeter wave devices; Mirrors; Phase measurement; Signal resolution; Stability; Temperature distribution;
  • fLanguage
    English
  • Journal_Title
    Instrumentation and Measurement, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9456
  • Type

    jour

  • DOI
    10.1109/TIM.1966.4313566
  • Filename
    4313566