DocumentCode :
1190659
Title :
Calibration and Stability of a Millimeter Confocal Wavemeter
Author :
Zamites, C.J., Jr.
Volume :
15
Issue :
4
fYear :
1966
Firstpage :
388
Lastpage :
392
Abstract :
The construction and calibration of a confocal wavemeter for the 94 GHz region is described. Material selection was predicated upon obtaining a frequency stability of 1 part in 105 per °C over a 34°C temperature range. The system uses a 94 GHz signal generated from a crystal reference and phase-lock loop to calibrate the wavemeter. Several readout techniques are discussed. 1) the conventional dip indication, 2) the pip obtained from the resonator used in the transmission mode, and 3) the measurement of the pip phase. The latter technique permits the resolution of 100 KHz at a center frequency of 100 GHz or 1 part in 106.
Keywords :
Calibration; Crystalline materials; Electromagnetic measurements; Frequency measurement; Millimeter wave devices; Mirrors; Phase measurement; Signal resolution; Stability; Temperature distribution;
fLanguage :
English
Journal_Title :
Instrumentation and Measurement, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9456
Type :
jour
DOI :
10.1109/TIM.1966.4313566
Filename :
4313566
Link To Document :
بازگشت