DocumentCode :
1191876
Title :
Dual slew-rate limiting for reducing gradient field interference in cardiac gated magnetic resonance imaging
Author :
Hughes, R.L.
Author_Institution :
Cardiac Magn. Resonance Unit, R. Brompton Hosp., London
Volume :
45
Issue :
6
fYear :
2009
Firstpage :
298
Lastpage :
300
Abstract :
A method is presented for minimising gradient field interference on the electrocardiogram (ECG) of a patient undergoing cardiac magnetic resonance imaging. A two lead ECG amplifier uses a single slew-rate limiter in the main signal path, this path providing the ECG signal plus slew-rate limited gradient interference from the scanner. A secondary signal path, also containing a slew-rate limiter, is derived from the main path and provides only slew-rate limited gradient interference. This signal can then be subtracted from the main path to cancel the original interference. This simple arrangement operates in real-time, needs no control signals from the scanner and has been successfully employed in a clinical environment.
Keywords :
biomedical MRI; electrocardiography; ECG; cardiac gated magnetic resonance imaging; dual slew rate; electrocardiogram; reducing gradient field interference; slew rate limited gradient interference;
fLanguage :
English
Journal_Title :
Electronics Letters
Publisher :
iet
ISSN :
0013-5194
Type :
jour
DOI :
10.1049/el.2009.3082
Filename :
4800377
Link To Document :
بازگشت