DocumentCode :
1192281
Title :
VHF and UHF Attenuation Measurements to 140 dB
Author :
Pulliam, S.Brown
Volume :
18
Issue :
1
fYear :
1969
fDate :
3/1/1969 12:00:00 AM
Firstpage :
7
Lastpage :
15
Abstract :
A system has been assembled for rapid comparison of production step attenuators, with a reference attenuator using direct substitution at spot frequencies in the VHF and UHF range. The total attenuation range for such tests is 140 dB, with a basic resolution of ±0.01 dB and with maximum errors of ±0.08 dB, attributable, primarily, to reflection inaccuracies. Simplified methods for measurement to 110 dB at any selected frequency are described, as well as additional refinements permitting spot frequency measurements to 170 dB. The basic equipment required for these measurements is a well-shielded signal source generating in excess of 100 mW, and a low-noise narrow-band receiver with provision for locking the signal frequency to the center of the receiver passband.
Keywords :
Assembly systems; Attenuation measurement; Attenuators; Frequency measurement; Narrowband; Production systems; Reflection; Signal generators; Testing; UHF measurements;
fLanguage :
English
Journal_Title :
Instrumentation and Measurement, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9456
Type :
jour
DOI :
10.1109/TIM.1969.4313753
Filename :
4313753
Link To Document :
بازگشت