Title :
A wideband millimeter wave six-port reflectometer using four diode detectors calibrated without a power ratio standard
Author :
Ghannouchi, Fadhel M. ; Bosisio, Renato G.
Author_Institution :
Microwave Res. Lab., Ecole Polytech. de Montreal, Que., Canada
fDate :
12/1/1991 12:00:00 AM
Abstract :
An 18-40 GHz coaxial six-port reflectometer using four diode detectors calibrated without a power ratio standard is presented. The hardware configuration of the six-port junction, the linearization of the diode detectors, and the calibration of the six-port reflectometer are discussed. Absolute power measurements or a high-quality variable attenuator is not required during the linearization of the diode detectors. A comparison between experimental results obtained by the six-port reflectometer and a Hewlett Packard HP 8510C Automatic Network Analyzer on a 3-dB attenuator terminated by a short circuit is presented
Keywords :
calibration; microwave reflectometry; reflectometers; 18 to 40 GHz; calibration; coaxial six port junction; diode detectors; linearization; wideband millimeter wave six-port reflectometer; Attenuators; Calibration; Circuits; Coaxial components; Diodes; Envelope detectors; Hardware; Millimeter wave measurements; Power measurement; Wideband;
Journal_Title :
Instrumentation and Measurement, IEEE Transactions on