DocumentCode :
1193708
Title :
Direct-Current Comparator Bridge for Resistance Thermometry
Author :
Kusters, Norbert L. ; MacMartin, Malcolm P.
Volume :
19
Issue :
4
fYear :
1970
Firstpage :
291
Lastpage :
297
Abstract :
The design, construction, and performance of a dc bridge that covers the complete range of both platinum-and germanium-resistance thermometry in four ranges, 0-11 up to 0-11 000 ohms, is described. The bridge uses a potentiometric method based on the current comparator. In the design, emphasis has been placed on reducing the noise level, on obtaining a single-balance operation, and on providing a recorder output. Detector sensitivity and noise level (with a three-second time constant) permit full eight-decade bridge resolution to be achieved at a thermometer current of I mA on all ranges except the lowest, which requires 5 mA. Bridge errors are believed to be less than 10-7 of a reading or one step on the last dial, whichever is larger. A facility is provided for measuring the self-heating effect of the thermometer, even if the measured temperature is not constant. The bridge is also suitable for the measurement of four-terminal standard resistors over the same ranges.
Keywords :
Bridge circuits; Councils; Detectors; Electrical resistance measurement; Impedance; Measurement standards; Noise level; Temperature measurement; Temperature sensors; Thermal resistance;
fLanguage :
English
Journal_Title :
Instrumentation and Measurement, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9456
Type :
jour
DOI :
10.1109/TIM.1970.4313916
Filename :
4313916
Link To Document :
بازگشت