• DocumentCode
    1194920
  • Title

    Basic evaluation of signal transmission coil in transcutaneous magnetic telemetry system for artificial hearts

  • Author

    Takura, T. ; Ishiai, H. ; Sato, F. ; Matsuki, H. ; Sato, T.

  • Author_Institution
    Dept. of Electr. & Commun. Eng., Graduate Sch. of Tohoku Univ., Miyagi, Japan
  • Volume
    41
  • Issue
    10
  • fYear
    2005
  • Firstpage
    4173
  • Lastpage
    4175
  • Abstract
    In implantable medical devices such as artificial hearts, it is important to communicate device control and monitoring signals across the skin for its stable operation. Using the signal transmission coil pairs we have proposed, we studied through measurements and experiments the best carrier frequency, based on the frequency characteristics of the L/C resonance circuit that inherits in the coil geometry.
  • Keywords
    artificial organs; bioelectric potentials; biomedical telemetry; coils; medical signal processing; L/C resonance circuit; artificial hearts; carrier frequency; coil geometry; device control; figure-eight coil; frequency characteristics; implantable medical devices; monitoring signals; signal transmission coil; transcutaneous magnetic telemetry system; transcutaneous signal transmission; Artificial heart; Biomedical monitoring; Coils; Frequency measurement; Implantable biomedical devices; Medical control systems; RLC circuits; Resonance; Skin; Telemetry; Carrier frequency; figure-eight coil; transcutaneous signal transmission;
  • fLanguage
    English
  • Journal_Title
    Magnetics, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9464
  • Type

    jour

  • DOI
    10.1109/TMAG.2005.854861
  • Filename
    1519571