DocumentCode :
1194920
Title :
Basic evaluation of signal transmission coil in transcutaneous magnetic telemetry system for artificial hearts
Author :
Takura, T. ; Ishiai, H. ; Sato, F. ; Matsuki, H. ; Sato, T.
Author_Institution :
Dept. of Electr. & Commun. Eng., Graduate Sch. of Tohoku Univ., Miyagi, Japan
Volume :
41
Issue :
10
fYear :
2005
Firstpage :
4173
Lastpage :
4175
Abstract :
In implantable medical devices such as artificial hearts, it is important to communicate device control and monitoring signals across the skin for its stable operation. Using the signal transmission coil pairs we have proposed, we studied through measurements and experiments the best carrier frequency, based on the frequency characteristics of the L/C resonance circuit that inherits in the coil geometry.
Keywords :
artificial organs; bioelectric potentials; biomedical telemetry; coils; medical signal processing; L/C resonance circuit; artificial hearts; carrier frequency; coil geometry; device control; figure-eight coil; frequency characteristics; implantable medical devices; monitoring signals; signal transmission coil; transcutaneous magnetic telemetry system; transcutaneous signal transmission; Artificial heart; Biomedical monitoring; Coils; Frequency measurement; Implantable biomedical devices; Medical control systems; RLC circuits; Resonance; Skin; Telemetry; Carrier frequency; figure-eight coil; transcutaneous signal transmission;