• DocumentCode
    1195198
  • Title

    Control of neuromuscular blockade in the presence of sensor faults

  • Author

    Lemos, João M. ; Magalhães, Hugo ; Mendonça, Teresa ; Dionisio, Rui

  • Author_Institution
    INESC-ID/IST, Lisboa, Portugal
  • Volume
    52
  • Issue
    11
  • fYear
    2005
  • Firstpage
    1902
  • Lastpage
    1911
  • Abstract
    The problem of embedding sensor fault tolerance in feedback control of neuromuscular blockade is considered. For tackling interruptions of feedback measurements, a structure based upon Bayesian inference as well as a predictive filter is proposed. This algorithm is general and can be applied to different situations. Here, it is incorporated in an adaptive automatic system for feedback control of neuromuscular blockade using continuous infusion of muscle relaxants. A significant contribution consists in the experimental clinical testing of the algorithm in patients undergoing surgery.
  • Keywords
    Bayes methods; biocontrol; feedback; median filters; medical signal processing; muscle; neurophysiology; sensors; surgery; Bayesian inference; feedback control; muscle relaxants; neuromuscular blockade; predictive filter; sensor fault tolerance; Adaptive control; Adaptive systems; Bayesian methods; Fault tolerance; Feedback control; Filters; Inference algorithms; Neurofeedback; Neuromuscular; Programmable control; Bayes inference; hybrid systems; median filter; neuromuscular blockade; outliers; physiological systems; supervisory multiple models control; Algorithms; Computer Simulation; Drug Therapy, Computer-Assisted; Electromyography; Feedback; Humans; Models, Neurological; Motor Neurons; Muscle, Skeletal; Neuromuscular Blockade; Neuromuscular Blocking Agents; Synaptic Transmission;
  • fLanguage
    English
  • Journal_Title
    Biomedical Engineering, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9294
  • Type

    jour

  • DOI
    10.1109/TBME.2005.856259
  • Filename
    1519599