Title :
The Reduction of Errors in a Precise Microwave Attenuator Calibration System
Author :
Hollway, D.L. ; Somlo, P.I.
Abstract :
Refinements have been made to the parallel-substitution heterodyne method of attenuation measurement. These are provision for small fixed frequency differences between the IF signals to improve the sensitivity of tests for leakage, a switched reference path to reduce errors from drift and to assess "flange scatter," a flexible waveguide in the measuring path to accommodate attenuators of different lengths, a steady flow of dry air through the system to remove moisture, and an electronic system for fine-balancing the two channels.
Keywords :
Attenuation measurement; Attenuators; Calibration; Electronic equipment testing; Flanges; Fluid flow measurement; Frequency measurement; Length measurement; Scattering; System testing;
Journal_Title :
Instrumentation and Measurement, IEEE Transactions on