DocumentCode :
1195949
Title :
Electromagnetic Interference Measurements of Fluorescent Lamps Operated with Solid-State Ballasts
Author :
Arthur, Allan A. ; Verderber, Rudolph R. ; Rubinstein, Francis ; Morse, Oliver C.
Author_Institution :
Lawrence Berkeley Laboratory, University of California, Berkeley, CA 94720.
Issue :
6
fYear :
1982
Firstpage :
647
Lastpage :
652
Abstract :
Solid-state ballasts were placed in fluorescent lamps in various areas of a hospital to determine if these high-frequency systems would adversely affect any hospital operations. The general areas tested included a lobby and an office space. Potentially sensitive areas containing hospital diagnostic and monitoring equipment, including a computerized axial tomography (CAT) scanning room, an electroencephalograph (EEG) examination room, and a coronary ward were also tested. The measurement techniques are described and the results discussed with respect to the existing radio-frequency interference (RFI) environment and with respect to electromagnetic interference (EMI) radiated and conducted limits specified by the Federal Communications Commission (FCC) and the Federal Drug Administration (FDA).
Keywords :
Computerized monitoring; Electromagnetic interference; Electromagnetic measurements; Electronic ballasts; FCC; Fluorescent lamps; Hospitals; Radiofrequency interference; Solid state circuits; Testing;