DocumentCode :
1197009
Title :
Standard Cell Calibration via Current Transfer
Author :
Williams, Edwin R. ; Olsen, Paul T. ; Field, Bruce F.
Volume :
23
Issue :
4
fYear :
1974
Firstpage :
299
Lastpage :
302
Abstract :
The EMF´s of standard cells are now being transferred between laboratories over a 1¿-km cable with a precision of 4 parts in 108 to provide an instantaneous comparison of the 2e/h and ¿p¿ experiments being carried out at the two facilities. This is accomplished by transferring a constant current that produces a 1-V drop across standard resistors located at both ends of the cable.
Keywords :
Calibration; Current measurement; Laboratories; Magnetic field measurement; Monitoring; NIST; Nuclear magnetic resonance; Protons; Resistors; Voltage;
fLanguage :
English
Journal_Title :
Instrumentation and Measurement, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9456
Type :
jour
DOI :
10.1109/TIM.1974.4314297
Filename :
4314297
Link To Document :
بازگشت