DocumentCode
1197462
Title
Corrections for Second-Order Doppler Shift and Cavity Phase Error in Cesium Atomic Beam Frequency Standards
Author
Daams, H.
Volume
23
Issue
4
fYear
1974
Firstpage
509
Lastpage
514
Abstract
From the measured static resonance curve of a cesium beam tube, corrections can be derived for second-order Doppler shift and cavity phase error. For the calculation Fourier transforms are used, which also produce the velocity distribution of the atoms. Data are given for the distortion of the resonance curve from which the corrections can be derived applicable in the case of very slow square wave frequency modulation. A method is also given for calculating corrections for faster modulation rates where the time-of-flight effects are no longer negligible. However, quadrature components will now occur in the beam signal, and their effect on the clock servo is discussed.
Keywords
Atomic beams; Atomic measurements; Clocks; Distortion measurement; Doppler shift; Error correction; Fourier transforms; Frequency modulation; Phase measurement; Resonance;
fLanguage
English
Journal_Title
Instrumentation and Measurement, IEEE Transactions on
Publisher
ieee
ISSN
0018-9456
Type
jour
DOI
10.1109/TIM.1974.4314342
Filename
4314342