Title :
A Voltage Doubling Circuit for the Absolute Calibration of 30-MHz Attenuation Measurement Systems
fDate :
3/1/1978 12:00:00 AM
Abstract :
The principle of Laverick´s microwave signal doubling method [1] has been adapted to 30 MHz, providing an absolute 6.0206-dB step to check the absolute accuracy of attenuation measuring systems operating at 30 MHz.
Keywords :
Attenuation measurement; Attenuators; Calibration; Couplers; Instruments; Measurement standards; Microwave measurements; Switches; Switching circuits; Voltage;
Journal_Title :
Instrumentation and Measurement, IEEE Transactions on
DOI :