DocumentCode :
1200005
Title :
A Voltage Doubling Circuit for the Absolute Calibration of 30-MHz Attenuation Measurement Systems
Author :
Somlo, P.I.
Volume :
27
Issue :
1
fYear :
1978
fDate :
3/1/1978 12:00:00 AM
Firstpage :
76
Lastpage :
79
Abstract :
The principle of Laverick´s microwave signal doubling method [1] has been adapted to 30 MHz, providing an absolute 6.0206-dB step to check the absolute accuracy of attenuation measuring systems operating at 30 MHz.
Keywords :
Attenuation measurement; Attenuators; Calibration; Couplers; Instruments; Measurement standards; Microwave measurements; Switches; Switching circuits; Voltage;
fLanguage :
English
Journal_Title :
Instrumentation and Measurement, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9456
Type :
jour
DOI :
10.1109/TIM.1978.4314623
Filename :
4314623
Link To Document :
بازگشت