NBS Provides Voltage Calibration Service in 0.1-10-Hz Range Using AC Voltmeter/Calibrator
Author
Schoenwetter, Howard K.
Volume
28
Issue
4
fYear
1979
Firstpage
327
Lastpage
331
Abstract
Prompted by the need to support vibration and pressure measurements at frequencies down to 0.5 Hz (with expected future needs to 0.1 Hz), NBS now offers a calibration service for voltage standards and rms voltmeters in the range of 0.1-10 Hz. The means for the service is an "ac Voltmeter/Calibrator," an NBS-developed instrument containing an rms digital voltmeter and ac and dc voltage calibrators. The methods used to calibrate the ac voltage calibrator are discussed; also, application of the ac Voltmeter/Calibrator to the calibration of customers\´ voltage and voltmeter standards is described. Finally, a multifrequency voltage reference source with frequency-independent amplitude is proposed as a more suitable transfer standard than thermal voltage converters (TVC\´s) for the 0.1-10-Hz range.
Keywords
Calibration; Frequency; Instruments; Measurement standards; NIST; Pressure measurement; Sea measurements; Vibration measurement; Voltage measurement; Voltmeters;