DocumentCode :
1202267
Title :
Capability and Limitations of Electrocardiography and Magnetocardiography
Author :
Plonsey, Robert
Author_Institution :
Department of Biomedical Engineering, Case Western Reserve University, Cleveland, Ohio 44106.
Issue :
3
fYear :
1972
fDate :
5/1/1972 12:00:00 AM
Firstpage :
239
Lastpage :
244
Abstract :
The capabilities and limitations of electrocardiography and magnetocardiography are discussed. Representing the electrical activity of the heart by an impressed current density ji, electrocardiography determines the spherical harmonic multipole expansion of its divergence (flux source), while magnetocardiography determines the spherical harmonic multipole expansion of the radial component of its curl (vortex source).
Keywords :
Biomedical measurements; Conductivity; Current density; Electrocardiography; Heart; Magnetic confinement; Magnetic field measurement; Magnetic flux; Torso; Voltage; Electrocardiography; Humans; Magnetics; Methods;