  • Abstract
    Methods are proposed for homogenizing and shimming a high homogeneity solenoid with a magnetic shield. The methods make the shield design more accurate and the shim coils less complicated. The whole-body 0.5 T superconducting MRI (magnetic resonance imaging) magnet is designed to achieve a field homogeneity of 3 ppm in a 50 cm DSV. Its fringe field is less than 0.5 mT outside a radius of 5 m using 12 tons of iron
  • Keywords
    biomedical NMR; biomedical equipment; magnetic fields; magnetic shielding; superconducting magnets; 0.5 T; MRI; fringe field; high homogeneity solenoid; homogeneous field; homogenizing; magnetic shield; shield design; shielded superconducting solenoid; shimming; Circuits; Geometry; Iron; Least squares methods; Magnetic analysis; Magnetic resonance imaging; Magnetic shielding; Solenoids; Superconducting coils; Superconducting magnets;