DocumentCode
1202458
Title
An Electronic Resuscitation Evaluation System
Author
Patrick, Herbert ; Eisenberg, Lawrence
Author_Institution
Emergency Care Research Institute, 913 Walnut Street, Philadelphia, Pa. 19107.
Issue
4
fYear
1972
fDate
7/1/1972 12:00:00 AM
Firstpage
317
Lastpage
320
Abstract
An electronic system is described that distinguishes and tabulates pressure pulses received from the Resusci-Anne training mannequin during cardiopulmonary resuscitation (CPR). The requirements of this system are in accordance with accepted CPR training programs. Samples of quantitative data gathered during mannequin practice sessions are included. The effectiveness of the system as a teaching aid is investigated, and correlations are made between CPR skills and a trainee´s physical attributes.
Keywords
Cardiac arrest; Cardiology; Cells (biology); Drugs; Education; Employment; Heart; Monitoring; Personnel; Ventilation;