• DocumentCode
    1202795
  • Title

    Computer-Controlled Automatic Comparison Measurements of Resistance Thermometers up to 464°C

  • Author

    Behnsch, Wolfgang ; Zinkernagel, Jorn

  • Volume
    29
  • Issue
    4
  • fYear
    1980
  • Firstpage
    334
  • Lastpage
    337
  • Abstract
    A computer-controlled test set is described which enables completely automatic measurements of resistance thermometers by comparison with a standard resistance thermometer. The automation involves heating control of a 3-zone oven, measurement data request, and data evaluation. The method of comparing measurements is to replace the "time consuming" temperature fixed point calibration procedure with those resistance thermometers the measurement accuracy of which is worse than 10 mK in the range of 0°C to 500°C. With appropriate resistance thermometers the extension of measurement range up to 1000°C is possible without any variation of the test set.
  • Keywords
    Automatic control; Automatic testing; Automation; Electrical resistance measurement; Heating; Measurement standards; Ovens; Temperature control; Thermal resistance; Time measurement;
  • fLanguage
    English
  • Journal_Title
    Instrumentation and Measurement, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9456
  • Type

    jour

  • DOI
    10.1109/TIM.1980.4314947
  • Filename
    4314947