• DocumentCode
    1202859
  • Title

    Automated Calibration of Blood Pressure Signal Conditioners

  • Author

    Martin, R. W. ; Weil, M. ; Shubin, H. ; Palley, N. ; Carrington, J. H. ; Bisera, J. ; Boycks, E. C.

  • Author_Institution
    Shock Research Unit and the Department of Medicine, University of Southern California School of Medicine, Los Angeles, Calif. 90027.
  • Issue
    1
  • fYear
    1973
  • Firstpage
    55
  • Lastpage
    58
  • Abstract
    A system for automatic calibration of blood pressure transducers is described. Automatic calibration is accomplished by activation of an analog-controlled servo system that sets the zero balance and gain in a single sequence. An unlimited number of transducers may be calibrated simultaneously, and calibration is completed within a period of less than 1 min. Accuracy and repeatability are assured without dependence on the operator´s judgment.