DocumentCode
1202961
Title
A Semiautomatic System for AC/DC Difference Calibration
Author
Lentner, Keith J. ; Childers, Clifton B. ; Tremaine, Susan G.
Volume
29
Issue
4
fYear
1980
Firstpage
400
Lastpage
405
Abstract
A semiautomatic ac/dc difference calibration system is described. The system operates over a frequency range of 20 Hz to 100 kHz, covering the voltage range from 0.5 V to 1 kV. For all voltages at frequencies in the range from 20 Hz to 20 kHz, the total uncertainty is 50 ppm and 100 ppm for voltages at frequencies between 20 kHz and 100 kHz. In addition to ac/dc difference testing, the system can be readily adapted to calibrate precision ac digital voltmeters or ac calibrators. Results of extensive intercomparison testing of the new system against a manual test system, using a multirange thermal transfer instrument as a transport standard, are reported, and the results indicate that the differences obtained are well within the combined total uncertainty limits of the two systems.
Keywords
Calibration; Circuit testing; Electronic equipment testing; Frequency; Instruments; NIST; System testing; Tellurium; Uncertainty; Voltage;
fLanguage
English
Journal_Title
Instrumentation and Measurement, IEEE Transactions on
Publisher
ieee
ISSN
0018-9456
Type
jour
DOI
10.1109/TIM.1980.4314966
Filename
4314966
Link To Document