• DocumentCode
    1203342
  • Title

    Mechanical Factors in the Design of Chronic Recording Intracortical Microelectrodes

  • Author

    Goldstein, Seth R. ; Salcman, Michael

  • Author_Institution
    Mechanical Engineering Section, Biomedical Engineering and Instrumentation Branch, National Institutes of Health, Bethesda, Md. 20014.
  • Issue
    4
  • fYear
    1973
  • fDate
    7/1/1973 12:00:00 AM
  • Firstpage
    260
  • Lastpage
    269
  • Abstract
    To obtain chronic recordings from single nerve cells in the cerebral cortex it is necessary to prevent motion of the microelectrode tip with respect to the adjacent neural tissue. This paper presents an engineering analysis to estimate these motions in terms of the geometry and material properties of the electrode, its lead wire, and the cortex. The results are used to gain an understanding of a recently reported successful recording system, and to thereby provide a design guide for future systems. In addition, a buckling analysis is presented for a class of microelectrodes to enable a minimum size to be achieved with an adequate safety margin.
  • Keywords
    Cerebral cortex; Electrodes; Geometry; Material properties; Mechanical factors; Microelectrodes; Motion analysis; Motion estimation; Safety; Wire; Animals; Biomedical Engineering; Electrodes, Implanted; Haplorhini; Mathematics; Microelectrodes; Motion;
  • fLanguage
    English
  • Journal_Title
    Biomedical Engineering, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9294
  • Type

    jour

  • DOI
    10.1109/TBME.1973.324190
  • Filename
    4120642