Pitfalls Using Doppler Ultrasound to Transduce Blood Flow
Author :
Flax, Stephen W. ; Webster, John G. ; Updike, Stuart J.
Author_Institution :
Department of Electrical Engineering, University of Wisconsin, Madison, Wis. 53706.
Issue :
4
fYear :
1973
fDate :
7/1/1973 12:00:00 AM
Firstpage :
306
Lastpage :
309
Abstract :
General testing procedures which will ascertain whether a Doppler ultrasonic blood flowmeter is correctly designed and operating properly are presented. A commercial Doppler ultrasonic blood flowmeter using the procedure is evaluated. Two pitfalls were encountered which prevented its proper operation. High-frequency response was inadequate, which introduced a nonlinearity into the input-output calibration curve. Low-frequency gain was too high, resulting in wall-motion artifacts. These problem areas are described in detail, and a remedy for each is suggested.
Keywords :
Blood flow; Disk recording; Frequency; Neurons; Oscilloscopes; Output feedback; Pulse amplifiers; Pulse circuits; Ultrasonic imaging; Voltage; Blood Flow Velocity; Transducers; Ultrasonics;