• DocumentCode
    1203570
  • Title

    Polarization and Corrosion Studies of Porous and Solid Anodes for Implantable Power-Generating Electrodes

  • Author

    DeRosa, Joseph F. ; Beard, Richard B. ; Carim, Hatim M. ; Dubin, Stephen E.

  • Author_Institution
    Biomedical Engineering and Science program, Drexel University, Philadelphia, Pa. 19104.
  • Issue
    5
  • fYear
    1973
  • Firstpage
    345
  • Lastpage
    349
  • Abstract
    An implantable power source for driving a pacemaker consisting of a hybrid cell utilizing a sacrificial porous aluminum anode and a catalytic cathode for reducing oxygen is considered. Porous aluminum anodes show a significant improvement in polarization over a solid aluminum anode under in vitro transient operating conditions found with certain types of pacemakers. However, the in vivo transient behavior of porous aluminum anodes is only slightly better than solid aluminum anodes. Using in vivo and in vitro linear polarization studies, the corrosion of solid aluminum has been determined to be more than an order of magnitude less than that of solid zinc. The weight loss of aluminum due to corrosion is less than 3.0 percent that of the aluminum needed to supply the faradaic current to the pacemaker. A hybrid cell consisting of an aluminum anode and a porous palladium-black cathode has powered a commercially available pacemaker. Further histopathologic evaluation of tissues surrounding anodes is needed before selecting either zinc or aluminum as an anode material.
  • Keywords
    Aluminum; Anodes; Cathodes; Corrosion; Electrodes; In vitro; In vivo; Pacemakers; Polarization; Solids; Aluminum; Animals; Chemistry, Physical; Corrosion; Dogs; Electricity; Electrodes, Implanted; Microscopy, Electron, Scanning; Pacemaker, Artificial; Surface Properties; Zinc;
  • fLanguage
    English
  • Journal_Title
    Biomedical Engineering, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9294
  • Type

    jour

  • DOI
    10.1109/TBME.1973.324286
  • Filename
    4120668