Title :
The Maintenance and Comparison of Standards of Electromotive Force at the BIPM
Author :
Witt, Thomas J. ; Reymann, Dominique
Volume :
32
Issue :
1
fYear :
1983
fDate :
3/1/1983 12:00:00 AM
Firstpage :
260
Lastpage :
266
Abstract :
This article describes improved standard cell comparison equipment, temperature-controlled enclosures, and measurement techniques used to relate reference groups of high-quality standard cells to the BIPM (Bureau International des Poids et Mesures) Josephson-effect standard with an uncertainty of less than 10 nV. This performance is demonstrated by results obtained over the last four years. The same techniques are now being applied in the routine comparison of international standards of electromotive force by the transfer of standard cells. An analysis of the behavior of the cells that have participated in comparisons over the last 12 years indicates that the uncertainty due to the instability of cells as transfer standards is about 0.5 ¿V, i.e., one hundred times greater than the best reproducibility in the laboratory. A comprehensive tabulation of uncertainties shows this to be by far the greatest single contribution.
Keywords :
Calibration; Force measurement; Frequency conversion; Histograms; Josephson effect; Laboratories; Measurement standards; Measurement techniques; Reproducibility of results; Uncertainty;
fLanguage :
English
Journal_Title :
Instrumentation and Measurement, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9456
Type :
jour
DOI :
10.1109/TIM.1983.4315055
Filename :
4315055