DocumentCode :
1204058
Title :
A Device for the Calculation of dP/dt/P with Internal Calibration
Author :
Brower, R.W. ; Roelandt, J. ; Meester, G.T.
Author_Institution :
Thoraxcentrum, Medical Faculty Rotterdam, Rotterdam, The Netherlands.
Issue :
1
fYear :
1974
Firstpage :
56
Lastpage :
60
Abstract :
An analog method for the calculation of dP/dt/P is described that in-corporates a means for internal calibration. Engineering tests of this device are reported as well as tests using pressures from the left ventricle of swine (20-25 Kgm). Interventions included inflation of a balloon in the inferior vena cava (peak pressure reduced by 50 percent), occlusion of the aorta (peak pressure increased by 100 percent), and pacing (from 150 to 210 beats/min). dP/dt/P (P denotes left-ventricular pressure) calculated from the analog device agreed within the noise level when compared with the calculation based on pressure alone using an A-D converter and digital computer.
Keywords :
Analog computers; Analog-digital conversion; Bandwidth; Calibration; Circuits; Filtering; Muscles; Myocardium; Noise level; Testing; Animals; Biomedical Engineering; Blood Pressure; Heart Function Tests; Heart Rate; Heart Ventricles; Pacemaker, Artificial; Swine; Vena Cava, Inferior;
fLanguage :
English
Journal_Title :
Biomedical Engineering, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9294
Type :
jour
DOI :
10.1109/TBME.1974.324365
Filename :
4120721
Link To Document :
بازگشت